The Blood Gas Analyzer measures pH, pCO2, pO2 and Hct from 95 µL of whole blood or 50 µL of capillary blood and reports results in under 90 seconds. From these it calculates temperature-corrected values, HCO3, TCO2, base excess, SBC, SO2%, anion gap, RI, AaDO2, O2SAT and calculated hemoglobin.
Calibration runs from a disposable cartridge that contains flush solution, mixed gas, liquid calibrants and the waste container, keeping operators away from biohazardous waste. The Blood Gas Analyzer has a 10.4-inch touch screen with self-diagnostic tutorials, an integrated barcode reader, a built-in 70 mm thermal printer and HL7 LIS connectivity.
Technical Specifications
undefined
- Brand: Master Science
- Model Number: MSBGA-95-4SWISS
- SKU: MSBGA-95-4SWISS
- Measured Parameters: pH, pCO2, pO2, Hct
- pH Range: 6.000-9.000
- pCO2 Range: 5.0-200.0 mmHg (0.667-26.67 kPa)
- pO2 Range: 0-800.0 mmHg (0-106.7 kPa)
- Hct Range: 12.0-65.0%
- Barometric Pressure: 500-800 mmHg
- Calculated Parameters: pH/pCO2/pO2 (TC), TCO2, HCO3, BEb, BEecf, SBC, SO2%, AG, RI, AaDO2, O2SAT, tHb(c)
- Sample Types: Whole blood, serum, plasma, dialysate, CSF
- Sample Volume: 95 µL whole blood; 50 µL capillary blood
- Analysis Time: < 90 s from aspiration; < 120 s sample to sample
- Input Parameters: Patient temperature, hemoglobin, FiO2, barometric pressure
- Calibration: Automatic, programmable 1-point and 2-point
- Calibrant Cartridge: Test volumes 300/200/150/100/50; 30 days on board; room-temperature storage
- Display: 10.4-inch TFT touch screen
- Data Storage: > 10,000 results
- Printer: Built-in 70 mm thermal printer
- Barcode Reader: Integrated, up to 14 barcode types
- Interfaces: RS-232, TCP/IP, USB; LIS under HL7
- Battery: Ni-MH, 3.0 Ah
- Dimensions (W × H × D): 400 × 574 × 344 mm
- Calibrant Pack Size (W × H × D): 258 × 180 × 70 mm
- Auto QC Cartridge Size (W × H × D): 103 × 180 × 70 mm
- Weight: 30 kg
- Operating / Storage Temperature: 15-30°C / 5-40°C
- Power Supply: 100-240 V, 50/60 Hz
Key Benefits
- Low Sample Volume: 95 µL of whole blood or 50 µL of capillary blood gives a full blood gas result.
- Closed Waste System: Waste is held inside the disposable calibrant cartridge.
- Fast Turnaround: Results in under 90 seconds and under 120 seconds between samples.
- Automated QC Option: The optional Auto QC cartridge runs three levels daily and stops testing after a QC failure.
Need electrolytes in the same run? See our Blood Gas and Electrolyte Analyzer, the extended version of the Blood Gas Analyzer.
Product Features
Touch Screen Interface
- 10.4-inch high-resolution touch screen
- Status area shows analyzer condition at a glance
- Online operation and self-diagnostic tutorials
- Infrared operator detection
Calibrant Cartridge
- Flush solution, mixed gas, liquid calibrants and waste in one pack
- Test volume options: 300, 200, 150, 100 or 50
- 30 days on board at room temperature
- Expiration alarm and remaining-test display
Measurement Unit
- Long-life, maintenance-free electrodes
- Preheats reagents and sample
- Adaptor for syringe and capillary samples
- Probe wiper for sample integrity and safety
Data Management
- Barcode input for sample, patient and consumables
- RS-232, TCP/IP and USB backup
- LIS connection under HL7
- Ni-MH battery, 3.0 Ah
Applications
undefined
- Arterial blood gas analysis in ICU and emergency rooms
- Acid-base balance assessment
- Oxygenation and ventilation monitoring
- Hematocrit checks during surgery
- Dialysate and CSF gas analysis
Standard Accessories Included
- Blood gas analyzer main unit
- Built-in thermal printer
- Integrated barcode reader
Optional Accessories
- Auto QC cartridge with three QC levels
- Calibrant cartridges in 300, 200, 150, 100 or 50 test sizes
Problems the Blood Gas Analyzer Solves
Critical-care teams face the issues below. The Blood Gas Analyzer solves each one.
- Large sample draws from fragile patients: Only 95 µL whole blood or 50 µL capillary blood is needed.
- Handling liquid biohazard waste: Waste stays sealed in the disposable cartridge.
- Slow results delay ventilation decisions: Results arrive in under 90 seconds.
- Electrode upkeep: Long-life electrodes need no maintenance.
- Manual QC scheduling: The optional Auto QC cartridge runs QC automatically.