The Master Science Swing Plasma Thawer is a digital constant-temperature thawing box that rocks the bags while they warm. The carrier moves left and right through 50 mm at 60 cycles per minute, and the temperature is held at 37 °C within ±0.1 °C.
It suits 10 bags per run and takes up to 20, thawing frozen plasma, platelets and other blood products in 10 to 15 minutes. Six custom modes, live temperature curves and stored run records are available on the IPS touch screen.
Technical Specifications
Specifications of the Master Science Swing Plasma Thawer:
- Brand: Master Science
- Model Number: MSPTM-1SKSWISS
- SKU: MSPTM-1SKSWISS
- Brand: Master Science
- Model Number: MSPTM-1SKSWISS
- SKU: MSPTM-1SKSWISS
- Thawing Method: Swing type
- Swing Frequency: 60 cycles/min
- Swing Mode and Amplitude: Left-right reciprocating, 50 mm ±5
- Water Storage Capacity: 48 kg ±5%
- Circulation Capacity: 30 L/min
- Temperature Control: 37 °C, adjustable
- Temperature Control Accuracy: ±0.1 °C
- Heating Power: 3500 W
- Heating System: PTC; room temperature to 37 °C in 10-12 min
- Suitable Thawing Quantity: 10 bags
- Maximum Thawing Quantity: 20 bags
- Thawing Time: 10 to 15 min
- Thawing Modes: 6 custom modes
- Display: IPS touch screen
- Data Export: USB
- External Dimensions (L×W×H): 550 × 500 × 945 mm
Key Benefits
- Bags kept in motion: the Master Science Swing Plasma Thawer reciprocates at 60 cycles per minute.
- Short cycle: 10 to 15 minutes per batch.
- Documented quality: temperature data stored and exportable.
- Simple upkeep: automatic filling, draining and cleaning.
- Safe by design: heater isolated from water, with over-temperature cut-off.
A non-swing version is the Master Science water bath plasma thawer.
Product Features
Temperature Control
- Live temperature and time curves
- Adjustable upper and lower limit alarms
- Over-temperature power-off and low water level protection
Records and Access
- Automatic storage of water and plasma temperature
- Hospital name and operator number saved with each record
- Password lock screen
- USB export
Water Management
- Automatic water intake and drainage
- Bags dried automatically after thawing
- One-touch cleaning with return to standby
- Detachable water tank
Use
- Adjustable thawing dividers
- Timed on/off
- Six user-defined thawing modes
Applications
The Master Science Swing Plasma Thawer is used for:
- Thawing frozen plasma
- Thawing platelets and blood products of different volumes
- Hospital blood banks and transfusion departments
Common Problems and How This Thawer Helps
How the Master Science Swing Plasma Thawer supports plasma thawing:
- Cold spots in static bags: left-right swing during thawing.
- Unrecorded thaw cycles: automatic logs with operator number.
- Wet bags after thawing: automatic drying step.
- Tank hygiene: removable tank and one-touch cleaning.