The Master Science Pulse Vacuum Steam Sterilizer is a large horizontal autoclave for hospital central supply departments, operating theatres, pharmaceutical plants and research facilities. It uses saturated steam, whose latent heat is released on condensation, to penetrate and sterilize textiles, surgical instruments, culture media and glassware, followed by vacuum drying.
A PLC with a touch-screen HMI runs every cycle automatically, records temperature data and monitors sterilization by both F0 value and temperature-time. With chambers from 150 L to 1500 L and a choice of single or double doors, the Master Science Pulse Vacuum Steam Sterilizer can be matched to anything from a clinic to a full CSSD.
Technical Specifications
Main technical parameters of the Master Science Pulse Vacuum Steam Sterilizer:
- Brand: Master Science
- Model Number: MSSTH-NS150DMSWISS, MSSTH-NS150SMSWISS, MSSTH-NS200DMSWISS, MSSTH-NS200SMSWISS, MSSTH-NS250DMSWISS, MSSTH-NS250SMSWISS, MSSTH-NS360DMSWISS, MSSTH-NS360SMSWISS, MSSTH-NS600DMSWISS, MSSTH-NS600SMSWISS, MSSTH-NS800DMSWISS, MSSTH-NS800SMSWISS, MSSTH-NS1000DMSWISS, MSSTH-NS1000SMSWISS, MSSTH-NS1200DMSWISS, MSSTH-NS1200SMSWISS, MSSTH-NS1500DMSWISS, MSSTH-NS1500SMSWISS
- SKU: MSSTH-NS150DMSWISS, MSSTH-NS150SMSWISS, MSSTH-NS200DMSWISS, MSSTH-NS200SMSWISS, MSSTH-NS250DMSWISS, MSSTH-NS250SMSWISS, MSSTH-NS360DMSWISS, MSSTH-NS360SMSWISS, MSSTH-NS600DMSWISS, MSSTH-NS600SMSWISS, MSSTH-NS800DMSWISS, MSSTH-NS800SMSWISS, MSSTH-NS1000DMSWISS, MSSTH-NS1000SMSWISS, MSSTH-NS1200DMSWISS, MSSTH-NS1200SMSWISS, MSSTH-NS1500DMSWISS, MSSTH-NS1500SMSWISS
- Brand: Master Science
- Models: MSSTH-NS[capacity]DMSWISS (single door) and MSSTH-NS[capacity]SMSWISS (double door), capacities 150, 200, 250, 360, 600, 800, 1000, 1200 and 1500 L
- SKU: MSSTH-NS150DMSWISS, MSSTH-NS150SMSWISS, MSSTH-NS200DMSWISS, MSSTH-NS200SMSWISS, MSSTH-NS250DMSWISS, MSSTH-NS250SMSWISS, MSSTH-NS360DMSWISS, MSSTH-NS360SMSWISS, MSSTH-NS600DMSWISS, MSSTH-NS600SMSWISS, MSSTH-NS800DMSWISS, MSSTH-NS800SMSWISS, MSSTH-NS1000DMSWISS, MSSTH-NS1000SMSWISS, MSSTH-NS1200DMSWISS, MSSTH-NS1200SMSWISS, MSSTH-NS1500DMSWISS, MSSTH-NS1500SMSWISS
- Chamber Dimensions (L × W × H): 150 L: 700 × 450 × 480 mm; 200 L: 925 × 450 × 480 mm; 250 L: 780 × 600 × 600 mm; 360 L: 1000 × 600 × 600 mm; 600 L: 1190 × 610 × 910 mm; 800 L: 1450 × 610 × 910 mm; 1000 L: 1250 × 680 × 1180 mm; 1200 L: 1450 × 680 × 1180 mm; 1500 L: 1850 × 680 × 1180 mm
- Design Pressure: -0.1 to 0.3 MPa
- Design Temperature: 150°C
- Working Pressure: 0.23 MPa
- Working Temperature: 115 to 138°C
- Vacuum Level: -0.086 MPa
- Temperature Uniformity: ≤ ±2.0°C
- Pulsing Cycles: 0 to 9 (adjustable)
- Water Source Pressure: 0.1 to 0.2 MPa
- Compressed Air Pressure: 0.4 to 0.7 MPa
- External Steam Pressure: 0.3 to 0.6 MPa
- Power Supply: 3-phase 5-wire, AC 380 V, 50 Hz
- Steam Generator: Built-in on 150-600 L models; external steam source required above 600 L
Key Benefits
- Deep steam penetration: Adjustable pulse vacuum removes air from packs, textiles and instruments before sterilization.
- Dual cycle assurance: F0 value and temperature-time monitoring together confirm each sterilization cycle.
- Clean/dirty separation: Optional double doors with interlock support separate loading and unloading areas.
- Controlled access: Three-level authority management limits who can change programmes.
For smaller jacketed units, see the horizontal jacketed autoclave 150L to 400L.
Product Features
Structure
- Horizontal rectangular double-layer chamber for safety, heat retention and drying
- Gear-engagement door with motorized locking
- Pneumatic door seal
- Single door or interlocked double door options
Piping and Validation
- High-quality stainless steel tubes
- Clamp-type quick-release fittings
- Standard equipment validation interfaces
Control System
- PLC with HMI touch screen
- F0 value and temperature-time monitoring
- Automatic temperature recording and complete cycle records
- Three-level authority management
Programmes
- Fabric
- Instruments
- Liquid
- BD test
- Pressure hold test
- Multiple custom programmes
Applications
The Master Science Pulse Vacuum Steam Sterilizer is used in:
- Hospital central sterile supply departments (CSSD)
- Operating room sterilization
- Pharmaceutical factories
- Research institutions and laboratories
- Sterilization of textiles, instruments, culture media and glassware
Standard Accessories Included
- Sterilizer main unit
- Sterilization racks (models below 600 L)
- Sterilization carts and transfer trolleys (600 L and above)
- Built-in steam generator (150-600 L models)
Optional Accessories
- Double-door configuration with interlock
Problems the Master Science Pulse Vacuum Steam Sterilizer Solves
High-volume sterilization departments face strict demands on capacity, documentation and safety. The Master Science Pulse Vacuum Steam Sterilizer helps with:
- Air pockets in dense loads: Repeated vacuum pulses ensure steam reaches every part of the load.
- Wet textiles: Vacuum drying leaves packs dry after the cycle.
- Incomplete records: Automatic temperature recording and F0 monitoring keep a full cycle history.
- Cross-contamination between areas: A double-door layout separates contaminated and sterile sides.