The Master Science Horizontal Steam Autoclave is a cylindrical, frame-mounted sterilizer for hospitals, clinics and laboratories that need to process large loads of dressings, instruments and media. Five chamber sizes are available, from 150 L up to 500 L, all with a stainless steel chamber and a radial-arm locking door.
The sterilizing course is controlled automatically and includes a drying stage, which makes the Master Science Horizontal Steam Autoclave well suited to drying dressings and linen after sterilization. Temperature can be set from 105 to 134°C, and both sterilization and drying times are adjustable from 0 to 60 minutes.
Technical Specifications
Technical data of the Master Science Horizontal Steam Autoclave:
- Brand: Master Science
- Model Number: MSSTH-150SWISS, MSSTH-200SWISS, MSSTH-280SWISS, MSSTH-400SWISS, MSSTH-500SWISS
- SKU: MSSTH-150SWISS, MSSTH-200SWISS, MSSTH-280SWISS, MSSTH-400SWISS, MSSTH-500SWISS
- Brand: Master Science
- Models: MSSTH-150SWISS (150 L), MSSTH-200SWISS (200 L), MSSTH-280SWISS (280 L), MSSTH-400SWISS (400 L), MSSTH-500SWISS (500 L)
- SKU: MSSTH-150SWISS, MSSTH-200SWISS, MSSTH-280SWISS, MSSTH-400SWISS, MSSTH-500SWISS
- Chamber Size: 150 L: Ø440 × 1000 mm; 200 L: Ø515 × 1000 mm; 280 L: Ø600 × 1000 mm; 400 L: Ø700 × 1100 mm; 500 L: Ø700 × 1300 mm
- Working Pressure: 0.22 MPa
- Working Temperature: 134°C
- Temperature Range: 105 to 134°C
- Sterilization Time: 0 to 60 min
- Drying Time: 0 to 60 min
- Heat Uniformity: ≤ ±2°C
- Power: 150/200 L: 9 kW; 280 L: 12 kW; 400/500 L: 15 kW; 380 V, 50/60 Hz
- External Dimensions: 150 L: 1350 × 600 × 1520 mm; 200 L: 1400 × 670 × 1650 mm; 280 L: 1400 × 770 × 1780 mm; 400 L: 1430 × 880 × 1830 mm; 500 L: 1800 × 900 × 1820 mm
Key Benefits
- Simple automatic operation: The sterilization course runs automatically once the programme is set.
- Dressing drying: A built-in drying stage removes moisture from dressings and fabric loads.
- Door safety: The door cannot be opened until chamber pressure falls to 0.027 MPa, and the cycle will not start with the door open.
- Wide size range: Five chamber volumes from 150 L to 500 L fit different workloads.
For F0 control and touch-screen records, compare the horizontal jacketed autoclave 150L to 400L.
Product Features
Construction
- Stainless steel sterilizing chamber
- Horizontal cylindrical design on a steel frame
- Radial-arm door locking wheel
Operation
- Automatically controlled sterilizing course
- Adjustable sterilization and drying times, 0 to 60 min
- Temperature range 105 to 134°C
Safety
- Over-temperature and over-pressure protection
- Door interlock until pressure drops to 0.027 MPa
- Safety valve opens above 0.24 MPa and exhausts steam to the water tank
- Automatic power cut-off and alarm on low water
Applications
The Master Science Horizontal Steam Autoclave is used for:
- Hospital and clinic sterilization departments
- Sterilizing and drying surgical dressings and linen
- Large-batch glassware and instrument sterilization
- Culture media sterilization in microbiology labs
- Waste decontamination in research facilities
Standard Accessories Included
Problems the Master Science Horizontal Steam Autoclave Solves
Facilities with high daily loads need a sterilizer that is robust and simple. The Master Science Horizontal Steam Autoclave helps with:
- Wet dressings after sterilization: The drying function removes moisture before unloading.
- Opening the door too early: A pressure interlock keeps the door closed until it is safe.
- Overpressure risk: The safety valve automatically vents steam above 0.24 MPa.
- Running dry: Low-water detection cuts power and sounds an alarm.