The ELISA Microplate Reader with Monochromator is a full-wavelength absorbance reader for 96-well and 384-well microplates. A grating monochromator selects any wavelength from 200 to 1000 nm in 1 nm steps, so ELISA, protein and nucleic acid assays can run without changing filters.
A xenon flash lamp and a double-beam design with two silicon photodetectors give a 0-4.0 OD measuring range with linearity of R² ≥ 0.999 from 0 to 3.0 Abs. The ELISA Microplate Reader with Monochromator is offered in two versions: the standard model and the HT model, which adds a cuvette slot for single-sample readings from 200 to 1000 nm with incubation.
Technical Specifications
undefined
- Brand: Master Science
- Model Number: MSMPR-A200SWISS,MSMPR-A200HTSWISS
- SKU: MSMPR-A200SWISS,MSMPR-A200HTSWISS
- Wavelength Selection: Monochromator
- Wavelength Range: 200-1000 nm, 1 nm step
- Light Source: Xenon flash lamp
- Wavelength Accuracy: 2 nm
- Wavelength Repeatability: 0.2 nm
- Bandwidth: < 2.5 nm
- Optical System: Double beam, 2 silicon photodetectors
- Plate Types: 96-well and 384-well plates
- Detection Mode: Absorbance
- Measuring Range: 0-4.0 OD
- Linearity: R² ≥ 0.999 (0-3.0 Abs)
- Absorbance Accuracy: ±(1.0% + 0.003 Abs) at 0-2.0 Abs; ±2.0% at 2.0-2.5 Abs
- Absorbance Repeatability: CV < 0.5% (accurate mode); CV < 1.0% (fast mode)
- Read Time (96-well): < 8 s (fast mode); < 28 s (accurate mode)
- Shaking: Linear, 3 speeds
- Incubation: RT+4°C to 45°C, ±0.5°C at 37°C
- Cuvette Slot (HT model): 200-1000 nm, with incubation
- Display: 10-inch touch screen
- Storage: 16 GB, more than 20,000 data files
- Interfaces: 1 × USB-B, 2 × USB-A, 1 × network port
- Power Supply: DC 24 V, 6.67 A
- Dimensions: 300 × 500 × 260 mm
- Net Weight: 15.5 kg
- Gross Weight: 21 kg
- Package Size: 690 × 490 × 500 mm
Key Benefits
- Any Wavelength, No Filters: The monochromator covers 200-1000 nm in 1 nm steps, so new assays need no extra filter purchases.
- Two Reading Modes: Fast mode reads a 96-well plate in under 8 s; accurate mode reads it in under 28 s with CV below 0.5%.
- Wide Linear Range: Double-beam optics give R² ≥ 0.999 from 0 to 3.0 Abs and a measuring range up to 4.0 OD.
- Built-In Plate Conditioning: Linear shaking at three speeds and incubation up to 45°C prepare samples inside the instrument.
For automated plate washing before reading, pair the ELISA Microplate Reader with Monochromator with our ELISA Microplate Washer.
Product Features
Optics
- Grating monochromator, 200-1000 nm in 1 nm increments
- Xenon flash lamp light source
- Wavelength accuracy 2 nm, repeatability 0.2 nm, bandwidth below 2.5 nm
- Double-beam path with two silicon photodetectors
Plate Handling
- Accepts 96-well and 384-well microplates
- Linear shaking with three speed settings
- Incubation from RT+4°C to 45°C, ±0.5°C at 37°C
Operation and Data
- 10-inch touch screen for stand-alone use
- 16 GB storage for more than 20,000 data files
- 1 USB-B, 2 USB-A and 1 network port
HT Model
- Additional cuvette slot for single samples
- Cuvette reading from 200 to 1000 nm
- Cuvette incubation
Applications
undefined
- ELISA and other enzyme immunoassays
- Protein quantification (BCA, Bradford, Lowry, UV 280 nm)
- Nucleic acid quantification at 260/280 nm
- Enzyme kinetics and activity assays
- Cell viability and proliferation assays (MTT, CCK-8)
- Wavelength scanning of unknown samples
Standard Accessories Included
- ELISA microplate reader main unit (standard or HT model)
- Power adapter (DC 24 V)
Optional Accessories
- PC software
- u-Nano ultra-micro plate (2-4 µL, 1-16 samples)
- ABS optical validation board
Problems the ELISA Microplate Reader with Monochromator Solves
Labs that rely on filter-based readers often meet the issues below. The ELISA Microplate Reader with Monochromator addresses each one.
- Filter readers only cover a few fixed wavelengths: The monochromator selects any wavelength from 200 to 1000 nm in 1 nm steps.
- Slow reading delays high-throughput work: Fast mode reads a full 96-well plate in under 8 seconds.
- Separate shaker and incubator take up bench space: Shaking at three speeds and incubation up to 45°C are built in.
- Single samples waste a whole plate: The HT model adds a cuvette slot for single-sample readings.
- Results lost without a PC: 16 GB internal storage holds more than 20,000 data files.