The Crude Fiber Analyzer measures raw (crude) fiber content in grain, feed, food and other agricultural and sideline products by the acid and alkali boiling method. Samples are boiled in acid and alkali of precise concentration under set conditions, soluble substances are removed with ether, and minerals are deducted by high-temperature ashing.
Six samples are processed per run, with a full test taking about 100 minutes: 40 minutes acid, 40 minutes alkali and 20 minutes filtration and washing. The Crude Fiber Analyzer achieves an absolute error of ≤ 0.4% for fiber below 10% and ≤ 4% above 10%.
Technical Specifications
undefined
- Brand: Master Science
- Model Number: MSFBA6SWISS
- SKU: MSFBA6SWISS
- Measuring Object: Crude fiber in grain, feed, food and agricultural products
- Method: Acid and alkali boiling
- Samples per Run: 6
- Reproducibility (< 10% fiber): Absolute error ≤ 0.4%
- Reproducibility (> 10% fiber): Absolute error ≤ 4%
- Measuring Time: About 100 min (acid 40 min, alkali 40 min, filter and wash 20 min)
- Water Supply Pressure: > 1.5 MPa
- Water Temperature: < 20°C
- Power Supply: AC 220 V, 50 Hz
- Power: 2800 W
- Dimensions: 670 × 450 × 670 mm
- Weight: 30 kg
Key Benefits
- Six Samples Together: Parallel columns process six samples per run.
- Standard Method: Acid and alkali boiling follows the classic crude fiber procedure.
- Predictable Timing: A full test takes about 100 minutes.
- Reliable Repeatability: Absolute error ≤ 0.4% for fiber below 10%.
For NDF, ADF and lignin testing, see our Infrared Fiber Analyzer, an advanced partner to the Crude Fiber Analyzer.
Product Features
Method
- Acid and alkali boiling
- Ether removal of soluble substances
- Mineral deduction by high-temperature burning
Capacity and Timing
- 6 samples per run
- Acid stage 40 min, alkali stage 40 min
- Filtration and washing 20 min
Controls
- Front-panel switches and meters
- Filtrate and recoil control knobs
- Water supply above 1.5 MPa, below 20°C
Applications
undefined
- Crude fiber testing in animal feed
- Fiber content in grain and cereals
- Food fiber analysis
- Agricultural product quality control
Standard Accessories Included
- Crude fiber analyzer main unit with 6 glass columns
Optional Accessories
- No optional accessories are listed for this model; contact our team for compatible items.
Problems the Crude Fiber Analyzer Solves
Beaker-based fiber digestion causes the issues below. The Crude Fiber Analyzer handles each one.
- One sample per beaker setup: Six samples run in parallel.
- Inconsistent boiling conditions: Each column follows the same controlled acid and alkali steps.
- Slow, scattered filtration: Integrated filtration and washing finish in about 20 minutes.
- Poor repeatability: Absolute error ≤ 0.4% for fiber below 10%.